Infosys在阅读器中打开 / AI 翻译

I/O Analog Manual | Notes on I/O components with analog in and outputs · 0 次阅读

Further considerations...

The following section contains some practical notes from everyday use of analog measurement technology:

  • Two aspects are particularly relevant when using signal generators as signal transducers, especially during initial measurement trials with analog terminals or during filter trials:
    • The amplitude of the output signal often drops suddenly with increasing frequency. For example, if 1 Vpp is set as the target amplitude, this is usually achieved cleanly for "slow" signals (less than 100 Hz, depending on the device), but not at frequencies greater than 100 Hz. This is then interpreted as an apparent measuring error of the analog input, since the analog input measures the real level. It is strongly recommended not to trust the graphical display of the signal generator but to measure the output signal with a third device, ideally a proven oscilloscope. It may be necessary to manually increase the amplitude at the signal generator, depending on the frequency, until the device reaches its control limit.
    • Signal generators with graphic display (screen) are easy to set up, but only provide a target output signal that may not necessarily correspond to the actual measurable signal. The impedance setting of the output and the ground reference can be partly responsible for this.
      Example: a Tektronix AFG3022B signal generator is connected on channel 1 in two-pole/differential mode to an ELM3004 in ±10 V mode and a signal with 1 Vpp, 1 Hz is set.
      Further considerations... 1:
      However, the terminal actually sees 2 Vpp, represented below by an associated TwinCAT Scope output:
      Further considerations... 2:
      The reason for the "wrong" display of the signal generator is its output setting of "50 Ω" or "Load", i.e. it assumes a power-based level adjustment and thus also 50 Ω on the input side.
      Further considerations... 3:
      However, industrial analog terminals usually have input resistances of several 100 kΩ to MΩ. Therefore the impedance setting "HighZ" is the correct one in this case:
      Further considerations... 4:
      Electrically, the output signal has not changed during the changeover! Only the display on the screen changes.
      Again, it is recommended to measure the output signal with a third device, a multimeter or oscilloscope, before starting the test. No dynamic signal [kHz] is required for this test, a DC or slow AC signal is sufficient.
  • The source impedance in relation to the measuring device impedance is of great importance for correct measurement! Increased source impedance (high output resistance) means that the source can only drive a small amount of current. This has implications in three areas:
    • In general, the effect of the unavoidable voltage divider RSource/RMeasuring device can be observed: the measurable signal amplitude sometimes changes considerably if overall “slow” changes occur at one point in the system (measuring source - measuring device): e.g. the device of one side is exchanged or the resistances change due to temperature (self-heating, daylight, ...).
    • Signals in the LF/HF range are affected by unavoidable capacitive/inductive cable loads: if necessary, an intended voltage amplitude is not achieved because the source “does not provide sufficient power” due to the high source impedance and the “low-pass effect” of the cable load on an alternating quantity becomes apparent in such a way that “too little time is available to build up the signal”.
    • Furthermore, “fast” changes to the system can eventually lead to irritation of the low-power source and to resonances (see also Note on oscillation effects with analog 20 mA inputs).
  • The previously mentioned effects can lead to a problem area when measuring small signal voltages, which would have to be considered for low-power sources such as thermocouples.
    Example: IR sensors (respond to temperature radiation, comparatively like thermocouples) are usually high impedance (some 10 kΩ) and low power. In connection with a multiplex input circuit (keyword: “fast” system change) and a standard thermocouple terminal, there may be a reaction on the connecting line, a measurement close to the application is hardly possible:
    Further considerations... 5: A simultaneous acquisition (e.g. EL3314-0002) or very high-impedance measuring devices (e.g. ELM3344) provide a remedy.